FDA Device recall Z-0981-2024

Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410

On 2024-02-14 the FDA classified a Class II recall of Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410 by Beckman Coulter, citing there is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and b….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0981-2024
ClassificationClass II
Statusongoing
Initiated2023-11-21
Published2024-02-14
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter
DistributionUS

Product

Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410

Reason

There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.

Identifiers

code_infoUDI/DI 15099590230623, Lot Numbers: 125533, 125589, 233656, 233919, 234055, 234339, 234388