RecallRadar

FDA Device recall Z-0979-2019

Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364

On 2019-03-20 the FDA classified a Class I recall of Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364 by Cook, citing the products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0979-2019
ClassificationClass I
Statusterminated
Initiated2019-02-01
Published2019-03-20
Last updated2026-09-13 11:55 UTC
CompanyCook
DistributionUS

Product

Cook Transseptal Needle, Reference Number TSNC-18-71.0, Order Number G02364. For transseptal left heart access in both diagnostic and interventional procedures.

Reason

The products were manufactured without a back bevel on the needle tip. This could cause damage to the inside of the introducer during needle insertion.

Identifiers

code_infoLot 8833687

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0979-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.