SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH, REF/Catalog Number L2KPP2, Siemens Material Number…

On 2017-01-18 the FDA classified a Class II recall of SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH, REF/Catalog Number L2KPP2, Siemens Material Number… by Siemens Healthcare Diagnostics, citing siemens is conducting this recall for the IMMULITE¿ 2000/IMMULITE¿ 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has confirmed that IMMULITE¿….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0979-2017
ClassificationClass II
Statusterminated
Initiated2016-11-22
Published2017-01-18
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH, REF/Catalog Number L2KPP2, Siemens Material Number (SMN) 10381441 and REF/Catalog Number L2KPP6, SMN 10381442; IVD --- Made in UK -- Siemens healthcare Diagnostics Products Ltd. Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL UK For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers - for the quantitative measurement of intact parathyroid hormone (parathyrin, PTH) in EDTA plasma and serum, as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

Reason

Siemens is conducting this recall for the IMMULITE¿ 2000/IMMULITE¿ 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has confirmed that IMMULITE¿ 2000/IMMULITE¿ 2000 XPi Intact PTH kit lot 320 can exhibit an average negative bias of up to -39% at iPTH concentrations <20 pg/mL with serum and EDTA patient samples vs. a reference kit lot.

Identifiers

code_infoLot Number 320, exp. date 2016/11/30; UDI #s: 0063041496191032020161130 and 0063041496192732020161130

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.