FDA Device recall Z-0977-2022

Artis Q zeego, Model Number 10848283

On 2022-04-27 the FDA classified a Class II recall of Artis Q zeego, Model Number 10848283 by Siemens Medical Solutions Usa, citing if the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system wil….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0977-2022
ClassificationClass II
Statusongoing
Initiated2022-02-23
Published2022-04-27
Last updated2026-09-13 11:56 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Artis Q zeego, Model Number 10848283

Reason

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Identifiers

code_info
UDI: 4056869010007; Serial Numbers: 117021 117136 117324 117108 117001 117023 117329 117315 117125 117006 117312 117124…UDI: 4056869010007; Serial Numbers: 117021 117136 117324 117108 117001 117023 117329 117315 117125 117006 117312 117124 117129 117112 117034 117026 117029 117123 117100 117307 117117 117111 117304 117024 117325 117303 117030 117314 117309 117113 117322 117014 117040 117041 117114 117009 Additional Serial Numbers as of 10/4/23: 117127 117310

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.