RecallRadar

FDA Device recall Z-0977-2019

e600 Automatic Transport Ventilator, Model Number 01EVE600

On 2019-04-10 the FDA classified a Class I recall of e600 Automatic Transport Ventilator, Model Number 01EVE600 by O Two Medical Technologies, citing there is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher tha….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0977-2019
ClassificationClass I
Statusterminated
Initiated2019-02-12
Published2019-04-10
Last updated2026-09-13 11:55 UTC
CompanyO Two Medical Technologies
DistributionAZ, CA, CO, FL, OH, PA, TX, VA

Product

e600 Automatic Transport Ventilator, Model Number 01EVE600

Reason

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Identifiers

code_infoEV60001 to EV60126

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0977-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.