RecallRadar

FDA Device recall Z-0974-2026

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software…

On 2025-12-31 the FDA classified a Class II recall of Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software… by Zimmer, citing affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture oc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0974-2026
ClassificationClass II
Statusongoing
Initiated2025-12-02
Published2025-12-31
Last updated2026-09-13 11:57 UTC
CompanyZimmer
DistributionUS

Product

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Left, Trochanteric Nail Component: N/A

Reason

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Identifiers

code_infoLot Code: Model No 815609300 ; UDI-DI (01)00887868144462(17)340826(10)66717600 ; Lot Number 66717600

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0974-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.