RecallRadar

FDA Device recall Z-0974-2024

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

On 2024-02-14 the FDA classified a Class II recall of Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system by Philips North America, citing magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0974-2024
ClassificationClass II
Statusongoing
Initiated2022-01-06
Published2024-02-14
Last updated2026-09-13 11:57 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Reason

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Identifiers

code_infoUDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0974-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.