FDA Device recall Z-0974-2020
Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
- FDA Device
- Class II
- terminated
- Published 2020-02-12
On 2020-02-12 the FDA classified a Class II recall of Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151 by Cook, citing the wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0974-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-01-04 |
| Published | 2020-02-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Cook |
| Distribution | US |
Product
Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151
Reason
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Identifiers
| code_info | 8011027 8183322 8037660 8030233 8018054 7994816 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0974-2020%22