RecallRadar

FDA Device recall Z-0972-2022

Artis zee biplane MN, Model Number 10094143

On 2022-04-27 the FDA classified a Class II recall of Artis zee biplane MN, Model Number 10094143 by Siemens Medical Solutions Usa, citing if the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system wil….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0972-2022
ClassificationClass II
Statusongoing
Initiated2022-02-23
Published2022-04-27
Last updated2026-09-13 11:56 UTC
CompanySiemens Medical Solutions Usa
DistributionUS

Product

Artis zee biplane MN, Model Number 10094143

Reason

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Identifiers

code_infoUDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0972-2022%22