FDA Device recall Z-0972-2022
Artis zee biplane MN, Model Number 10094143
- FDA Device
- Class II
- ongoing
- Published 2022-04-27
On 2022-04-27 the FDA classified a Class II recall of Artis zee biplane MN, Model Number 10094143 by Siemens Medical Solutions Usa, citing if the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system wil….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0972-2022 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2022-02-23 |
| Published | 2022-04-27 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
Artis zee biplane MN, Model Number 10094143
Reason
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Identifiers
| code_info | UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0972-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.