FDA Device recall Z-0968-2018
BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841
- FDA Device
- Class II
- terminated
- Published 2018-03-21
On 2018-03-21 the FDA classified a Class II recall of BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841 by Becton Dickinson And, citing A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specime….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0968-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-08-29 |
| Published | 2018-03-21 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Becton Dickinson And |
| Distribution | US |
Product
BD Vacutainer¿ K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.
Reason
A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.
Identifiers
| code_info | Catalog No. 367841 Lot No. 6279849 UDI: (01)30382903678410 |
|---|