FDA Device recall Z-0965-2023
CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f)…
- FDA Device
- Class I
- ongoing
- Published 2023-02-08
On 2023-02-08 the FDA classified a Class I recall of CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f)… by Smiths Medical Asd, citing there are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (ND….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0965-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2022-12-09 |
| Published | 2023-02-08 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Smiths Medical Asd |
| Distribution | US |
Product
CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device
Reason
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Identifiers
| code_info | a) REF 21-7600-24, UDI/DI 30610586044001, Lot Numbers: 4084914 through 4329633; b) REF 21-7609-24, UDI/DI 3061058604…a) REF 21-7600-24, UDI/DI 30610586044001, Lot Numbers: 4084914 through 4329633; b) REF 21-7609-24, UDI/DI 30610586044018, Lot Numbers: 4072200 through 4334090; c) REF 21-7624-24, UDI/DI 30610586044742, Lot Numbers: 4092506 through 4309481; d) REF 21-7649-24, UDI/DI 35019517184246, Lot Numbers: 4076410 through 4334357; e) REF 21-7600-24JP, UDI/DI N/A, Lot Numbers: 4084914 through 4329633, f) REF 21-7609-24JP, UDI/DI N/A, Lot Numbers: 4072200 through 4334090; g) REF 21-7624-24JP, UDI/DI N/A, Lot Numbers: 4092506 through 4309481; h) REF 21-7649-24JP, UDI/DI N/A, Lot Numbers: 4076410 through 4334357. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0965-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.