FDA Device recall Z-0965-2023

CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f)…

On 2023-02-08 the FDA classified a Class I recall of CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f)… by Smiths Medical Asd, citing there are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (ND….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0965-2023
ClassificationClass I
Statusongoing
Initiated2022-12-09
Published2023-02-08
Last updated2026-09-13 11:56 UTC
CompanySmiths Medical Asd
DistributionUS

Product

CADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-24JP, h) REF 21-7649-24JP Computerized Ambulatory Delivery Device

Reason

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Identifiers

code_info
a) REF 21-7600-24, UDI/DI 30610586044001, Lot Numbers: 4084914 through 4329633; b) REF 21-7609-24, UDI/DI 3061058604…a) REF 21-7600-24, UDI/DI 30610586044001, Lot Numbers: 4084914 through 4329633; b) REF 21-7609-24, UDI/DI 30610586044018, Lot Numbers: 4072200 through 4334090; c) REF 21-7624-24, UDI/DI 30610586044742, Lot Numbers: 4092506 through 4309481; d) REF 21-7649-24, UDI/DI 35019517184246, Lot Numbers: 4076410 through 4334357; e) REF 21-7600-24JP, UDI/DI N/A, Lot Numbers: 4084914 through 4329633, f) REF 21-7609-24JP, UDI/DI N/A, Lot Numbers: 4072200 through 4334090; g) REF 21-7624-24JP, UDI/DI N/A, Lot Numbers: 4092506 through 4309481; h) REF 21-7649-24JP, UDI/DI N/A, Lot Numbers: 4076410 through 4334357.

Status history