RecallRadar

FDA Device recall Z-0964-2023

CADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. C…

On 2023-02-08 the FDA classified a Class I recall of CADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. C… by Smiths Medical Asd, citing there are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (ND….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0964-2023
ClassificationClass I
Statusongoing
Initiated2022-12-09
Published2023-02-08
Last updated2026-09-13 11:56 UTC
CompanySmiths Medical Asd
DistributionUS

Product

CADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. Computerized Ambulatory Delivery Device

Reason

There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Identifiers

code_info
a) REF 21-7308-24, UDI/DI 30610586032374, Lot Numbers: 4053922 through 4334076; b) REF 21-7309-24, UDI/DI 30610586032…a) REF 21-7308-24, UDI/DI 30610586032374, Lot Numbers: 4053922 through 4334076; b) REF 21-7309-24, UDI/DI 30610586032381, Lot Numbers: 4062405 through 4334085; c) REF 21-7310-24, UDI/DI 30610586032398, Lot Numbers: 4062404 through 4330874 d) REF 21-7308-24JP, UDI/DI N/A, Lot Numbers: 4053922 through 4334075; e) REF 21-7309-24JP, UDI/DI N/A, Lot Numbers: 4062405 through 4330870.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0964-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.