FDA Device recall Z-0964-2020
Wire Guide, Guidewire, Global Product Number G14326
- FDA Device
- Class II
- terminated
- Published 2020-02-12
On 2020-02-12 the FDA classified a Class II recall of Wire Guide, Guidewire, Global Product Number G14326 by Cook, citing the wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0964-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-01-04 |
| Published | 2020-02-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Cook |
| Distribution | US |
Product
Wire Guide, Guidewire, Global Product Number G14326
Reason
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Identifiers
| code_info | 7857489 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0964-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.