FDA Device recall Z-0963-2026
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emi…
- FDA Device
- Class II
- ongoing
- Published 2025-12-31
On 2025-12-31 the FDA classified a Class II recall of The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emi… by Howmedica Osteonics, citing when the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0963-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-12 |
| Published | 2025-12-31 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Howmedica Osteonics |
| Distribution | US |
Product
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Reason
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Identifiers
| code_info | Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0963-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.