Fixed Core Straight Safety Wire Guide, Guidewire, Global Product Number G14285

On 2020-02-12 the FDA classified a Class II recall of Fixed Core Straight Safety Wire Guide, Guidewire, Global Product Number G14285 by Cook, citing the wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0960-2020
ClassificationClass II
Statusterminated
Initiated2019-01-04
Published2020-02-12
Last updated2026-09-13 11:55 UTC
CompanyCook
DistributionUS

Product

Fixed Core Straight Safety Wire Guide, Guidewire, Global Product Number G14285

Reason

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Identifiers

code_info7971639

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.