FDA Device recall Z-0957-2020
Medtronic MiniMed 620G Insulin Infusion Pump, Ref #s/Model(s) - MMT-1510, MMT-1710, and MMT-1750 *Not Distributed within the US
- FDA Device
- Class I
- ongoing
- Published 2020-02-19
On 2020-02-19 the FDA classified a Class I recall of Medtronic MiniMed 620G Insulin Infusion Pump, Ref #s/Model(s) - MMT-1510, MMT-1710, and MMT-1750 *Not Distributed within the US by Medtronic, citing there have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that pr….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0957-2020 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2019-11-21 |
| Published | 2020-02-19 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Medtronic |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
Product
Medtronic MiniMed 620G Insulin Infusion Pump, Ref #s/Model(s) - MMT-1510, MMT-1710, and MMT-1750 *Not Distributed within the US
Reason
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
Identifiers
| code_info | All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US*** |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0957-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.