RecallRadar

FDA Device recall Z-0952-2023

cobas infinity central lab, Material Number 07154003001

On 2023-01-18 the FDA classified a Class II recall of cobas infinity central lab, Material Number 07154003001 by Roche Diagnostics Operations, citing A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order recei….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0952-2023
ClassificationClass II
Statusongoing
Initiated2022-12-12
Published2023-01-18
Last updated2026-09-13 11:56 UTC
CompanyRoche Diagnostics Operations
DistributionUS

Product

cobas infinity central lab, Material Number 07154003001

Reason

A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions.

Identifiers

code_info
cobas infinity central lab / cobas infinity core license using software versions (2.5.x , 3.01.x, 3.02.x, and 3.03.x) t…cobas infinity central lab / cobas infinity core license using software versions (2.5.x , 3.01.x, 3.02.x, and 3.03.x) that have been configured with all of the following pre-conditions will be impacted: 1. Having Order ID configuration to include the date (Not Common in U.S. Configurations) 2. Having a host configured by Order rather than Tube (Not Common in U.S..Configurations) 3. Having message configured to receive registration date of the order using ExtRegisterDate and ExtRegisterHour. 4. Having HCA configuration enabled for " Reject orders if not possible to assign External ID" and the radio button option set to "To at least one test" UDI-DI 04015630936007