FDA Device recall Z-0952-2019
AF w/AS 550 x 0.5-7 ml,w/Y site,DL CPNB, Product code MVAXL2Y CPNB, Arrow AutoFuser Disposable Pain Control Pump
- FDA Device
- Class II
- terminated
- Published 2019-03-06
On 2019-03-06 the FDA classified a Class II recall of AF w/AS 550 x 0.5-7 ml,w/Y site,DL CPNB, Product code MVAXL2Y CPNB, Arrow AutoFuser Disposable Pain Control Pump by Arrow International, citing possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alt….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0952-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-10-02 |
| Published | 2019-03-06 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Arrow International |
| Distribution | US |
Product
AF w/AS 550 x 0.5-7 ml,w/Y site,DL CPNB, Product code MVAXL2Y CPNB, Arrow AutoFuser Disposable Pain Control Pump
Reason
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Identifiers
| code_info | 74A1600493 74E1701689 74A1701278 74F1601947 74B1700147 74G1601342 74B1700922 74H1601292 74C1601403 74H1602005 74C1…74A1600493 74E1701689 74A1701278 74F1601947 74B1700147 74G1601342 74B1700922 74H1601292 74C1601403 74H1602005 74C1601458 74J1601528 74C1601839 74K1601040 74C1602928 74K1601957 74C1702397 74L1601996 74D1600766 74L1602670 74E1600739 74M1501120 74E1601599 74M1600311 74E1700481 74M1600671 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0952-2019%22