FDA Device recall Z-0952-2013

Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006 - Non-Sterile

On 2013-03-27 the FDA classified a Class II recall of Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006 - Non-Sterile by Extremity Medical, citing the 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected sc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0952-2013
ClassificationClass II
Statusterminated
Initiated2013-01-10
Published2013-03-27
Last updated2026-09-13 11:53 UTC
CompanyExtremity Medical
DistributionAZ, CA, FL, IL, MD, NC, NH, NY, OH, OK, PA, TN, TX, VA

Product

Extremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006 - Non-Sterile. Contained in Extremity Medical Surgical Kits including CompressX, Hallu.X, IO Fix and XMCP. Used in the selection of screw size when used with appropriate depth gauge.

Reason

The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.

Identifiers

code_infoCatalog Number 101-00006; Lot Number 00921538.