RecallRadar

FDA Device recall Z-0951-2017

Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hosp…

On 2017-01-18 the FDA classified a Class II recall of Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hosp… by Natus Manufacturing, citing the fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0951-2017
ClassificationClass II
Statusterminated
Initiated2016-10-15
Published2017-01-18
Last updated2026-09-13 11:54 UTC
CompanyNatus Manufacturing
DistributionCA, FL, IL, IN, MA, MS, NC, NJ, NY, TX, VA, WA

Product

Natus Warmette Blanket Cabinet and Warmette Dual Cabinet The Natus Warmette Blanket Cabinet is designed to store and warm blankets and hospital linens; while, the Natus Warmette Dual Cabinet is designed to store and warm blankets and hospital linens in one compartment and fluids (irrigation and/or injection) in the other compartment within the storage times and warming temperatures recommended by the manufacturers of those items

Reason

The fuse blows during initial power up or after a period of time. This results in a complete loss of power to the Cabinet and in some cases has resulted in the fuse cover for the power inlet melting slightly.

Identifiers

code_info
Warmette Blanket Cabinet Serial Numbers: 10216709, 10216711, 10218559, 10219214, 10219337, 10219338, 10219339, 102193…Warmette Blanket Cabinet Serial Numbers: 10216709, 10216711, 10218559, 10219214, 10219337, 10219338, 10219339, 10219340, 10219387, 10219388, 10219869, 10219870, 10219871, 10219872, 10219873, 1028880. Warmette Dual Cabinet Serial Numbers: 10216713, 10216714, 10218500, 10218501, 10218522, 10218526, 10218528, 10218982, 10219257, 10219389, 10219390, 10220534, 10220536, 10220537, 10220538.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0951-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.