RecallRadar

FDA Device recall Z-0950-2019

AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

On 2019-03-06 the FDA classified a Class II recall of AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump by Arrow International, citing possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alt….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0950-2019
ClassificationClass II
Statusterminated
Initiated2018-10-02
Published2019-03-06
Last updated2026-09-13 11:55 UTC
CompanyArrow International
DistributionUS

Product

AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Reason

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Identifiers

code_info74B1602449

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0950-2019%22