FDA Device recall Z-0950-2015

Labeled in part: I Color Complete / Color Lens, blister packing, 14.5mm. (Additional colors include: 1) Illusion style, Chestnut Brown, Vio…

On 2015-01-21 the FDA classified a Class II recall of Labeled in part: I Color Complete / Color Lens, blister packing, 14.5mm. (Additional colors include: 1) Illusion style, Chestnut Brown, Vio… by Intertrade Imports, citing A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses. I Color Color Complete brand color contact lenses are u….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0950-2015
ClassificationClass II
Statusterminated
Initiated2014-09-10
Published2015-01-21
Last updated2026-09-13 11:53 UTC
CompanyIntertrade Imports
DistributionCA, FL, TX

Product

Labeled in part: I Color Complete / Color Lens, blister packing, 14.5mm. (Additional colors include: 1) Illusion style, Chestnut Brown, Violet and others and 2) 2 Tone style: like Misty Blue among others)Eye contact lens Product Usage - Eye contact lens

Reason

A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses. I Color Color Complete brand color contact lenses are unapproved medical devices.

Identifiers

code_infoALL LOTS / CODES

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0950-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.