RecallRadar

FDA Device recall Z-0949-2023

IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) f…

On 2023-01-18 the FDA classified a Class II recall of IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) f… by Siemens Healthcare Diagnostics, citing potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control result….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0949-2023
ClassificationClass II
Statusongoing
Initiated2022-12-07
Published2023-01-18
Last updated2026-09-13 11:56 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # L2KTY2(D) Siemens Material Number (SMN): 10381647

Reason

Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples

Identifiers

code_infoUDI: 00630414972084 Kit Lots: D431, D433, D434, D435, and D436

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.