RecallRadar

FDA Device recall Z-0949-2019

AF 550 x 5ML/HR CPNB, Product code MC0050XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

On 2019-03-06 the FDA classified a Class II recall of AF 550 x 5ML/HR CPNB, Product code MC0050XL CPNB, Arrow AutoFuser Disposable Pain Control Pump by Arrow International, citing possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alt….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0949-2019
ClassificationClass II
Statusterminated
Initiated2018-10-02
Published2019-03-06
Last updated2026-09-13 11:55 UTC
CompanyArrow International
DistributionUS

Product

AF 550 x 5ML/HR CPNB, Product code MC0050XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Reason

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Identifiers

code_info74C1601434 74E1602096 74C1602395 74F1602444 74C1703028 74F1700186 74D1700428 74G1702048 74D1701032

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0949-2019%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.