FDA Device recall Z-0947-2015
Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003)
- FDA Device
- Class II
- terminated
- Published 2015-01-21
On 2015-01-21 the FDA classified a Class II recall of Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003) by Phadia Us, citing during an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mism….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0947-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-20 |
| Published | 2015-01-21 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Phadia Us |
| Distribution | US |
Product
Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003); Multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments. Software utilized in multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments.
Reason
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.
Identifiers
| code_info | Part Number 12-3800-01 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.