FDA Device recall Z-0946-2015

Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO…

On 2015-01-14 the FDA classified a Class II recall of Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO… by Siemens Healthcare Diagnostics, citing dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0946-2015
ClassificationClass II
Statusterminated
Initiated2014-11-24
Published2015-01-14
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista Myoglobin Calibrator (MYO CAL), Lot 4FD085 Product usage: in vitro diagnostic product for the calibration of Myoglobin (MYO) method on the Dimension Vista System.

Reason

Dimension Vista Myoglobin Calibrator (MYO CAL) may produce a positive shift in MYO QC and Patient test results that exceeds acceptance criteria for this product. A positive shift observed up to 12% at MYO concentrations within and above the reference range of the assay. Depending on quality control limits, this drift may not be detected.

Identifiers

code_infoLot 4FD085, exp. 12-01-2014

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0946-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.