AF 550 x 5ML/H 6" CATH, 4.5" IN, Product code MC0050XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump

On 2019-03-06 the FDA classified a Class II recall of AF 550 x 5ML/H 6" CATH, 4.5" IN, Product code MC0050XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump by Arrow International, citing possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alt….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0945-2019
ClassificationClass II
Statusterminated
Initiated2018-10-02
Published2019-03-06
Last updated2026-09-13 11:55 UTC
CompanyArrow International
DistributionUS

Product

AF 550 x 5ML/H 6" CATH, 4.5" IN, Product code MC0050XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump

Reason

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Identifiers

code_info74J1601025

Status history