FDA Device recall Z-0941-2017
G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06, 9180-07, 9180-17, 9101-00, 9101-06 The Opticage Expand…
- FDA Device
- Class II
- terminated
- Published 2017-01-11
On 2017-01-11 the FDA classified a Class II recall of G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06, 9180-07, 9180-17, 9101-00, 9101-06 The Opticage Expand… by Interventional Spine, citing during implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device f….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0941-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-11-22 |
| Published | 2017-01-11 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Interventional Spine |
| Distribution | US |
Product
G3 Opticage Expandable Interbody Fusion Device, Model No. 9180-00, 9180-01, 9180-06, 9180-07, 9180-17, 9101-00, 9101-06 The Opticage Expandable Interbody Fusion Device is a posterior lumbar intervertebral body fusion device and is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Opticage Interbody Fusion Device can be implanted via posterior, transforaminal or lateral approach. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Candidates for surgery should be skeletally mature and have had six months of conservative treatment. These patients may have had primary or secondary surgery, but no previous fusion at the involved levels. The device is not intended to be used as a stand-alone device. It must be used with supplemental internal spinal fixation systems that have been cleared for use in the lumbar spine (i.e. facet screw fixation systems, facet compression devices and posterior pedicle screw and rod systems).
Reason
During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device from expanding.
Identifiers
| code_info | Lot no. 011416-D 021216-A 021216-B 032015-B 032015-C 033116-B 040915-B 040915-E 041416-A 060116-A 060116-B…Lot no. 011416-D 021216-A 021216-B 032015-B 032015-C 033116-B 040915-B 040915-E 041416-A 060116-A 060116-B 060116-C 061616-B 061915-A 061915-B 070714-D 071116-A 071116-B 071116-C 071516-A 071516-B 071916-A 081016-A 081016-B 081016-C 083116-B 083116-B 091516-A 112515-A 033116-C 030716-A 081016-E 090216-C 090216-C 090216-D 030716-B 011416-E 021216-C 081016-F 081016-G 090216-A 090216-A 090216-B 081016-H 083116-C 083116-C 083116-D |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0941-2017%22