FDA Device recall Z-0940-2026
MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
- FDA Device
- Class II
- ongoing
- Published 2025-12-24
On 2025-12-24 the FDA classified a Class II recall of MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) by Path, citing following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0940-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-11-06 |
| Published | 2025-12-24 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Path |
| Distribution | WI |
Product
MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
Reason
Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
Identifiers
| code_info | Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945,…Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Other recalls by Path
- SENTIERO SCREENING, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S
FDA Device · Class II · 2023-12-06 · Path - SENTIERO, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098
FDA Device · Class II · 2023-12-06 · Path - SENTIERO Screening2, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S2
FDA Device · Class II · 2023-12-06 · Path - AuDX PRO, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-AUP
FDA Device · Class II · 2023-12-06 · Path