FDA Device recall Z-0940-2015
IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000
- FDA Device
- Class II
- terminated
- Published 2015-01-14
On 2015-01-14 the FDA classified a Class II recall of IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000 by Intuitive Surgical, citing intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wh….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0940-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-12-05 |
| Published | 2015-01-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
IS3000 Vision Side Cart (VSC) used in conjunction with the da Vinci Si Surgical System IS3000. Product number 371683-05/Model Number VS3000 The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and noncardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.
Reason
Intuitive Surgical is initiating a voluntary correction related to da Vinci Si System (IS3000) Vision Side Carts, with threaded-in caster wheels, which can result in the caster wheel bolts gradually unthreading and the wheels loosening.
Identifiers
| code_info | Product number 371683-05/Model Number VS3000 Serial numbers - USA: SH0018, SH0057, SH0097, SH0132, SH0165, SH0235, S…Product number 371683-05/Model Number VS3000 Serial numbers - USA: SH0018, SH0057, SH0097, SH0132, SH0165, SH0235, SH0019, SH0058, SH0098, SH0133, SH0167, SH0248, SH0021, SH0059, SH0099, SH0134, SH0168, USG015, SH0024, SH0060, SH0100, SH0135, SH0169, USG072, SH0025,SH0064, SH0101, SH0136, SH0170, USG146, SH0026, SH0065, SH0102, SH0137,SH0171, USG213, SH0029, SH0066, SH0103, SH0138, SH0172, USG317, SH0030, SH0067, SH0104, SH0139, SH0175, USG352, SH0031, SH0068, SH0105, SH0140, SH0177, USG353, SH0032, SH0069, SH0106 ,SH0141, SH0179 ,USG445, SH0033, SH0070, SH0107, SH0142, SH0180, USG519, SH0035, SH0071, SH0109, SH0143, SH0182, USG559, SH0036, SH0072, SH0110, SH0144, SH0184, USG631,SH0038, SH0073, SH0112, SH0145, SH0185, USG700, SH0039, SH0076, SH0113, SH0146, SH0186, USG701,SH0040, SH0077, SH0115, SH0147, SH0187, USG702, SH0041, SH0078, SH0117, SH0149, SH0188 ,USG703, SH0042, SH0079, SH0118, SH0150, SH0189, USG721, SH0044, SH0080, SH0119, SH0151, SH0190 ,USG728, SH0045, SH0081, S |
|---|