RecallRadar

FDA Device recall Z-0938-2014

Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109

On 2014-02-19 the FDA classified a Class II recall of Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109 by Integra Lifesciences, citing integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0938-2014
ClassificationClass II
Statusterminated
Initiated2014-01-22
Published2014-02-19
Last updated2026-09-13 11:53 UTC
CompanyIntegra Lifesciences
DistributionAZ, CA, FL, IA, IL, LA, MD, NC, NM, TX, VA

Product

Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; Malibu Polyaxial Screwdriver, Standard, Model Number 91-2110. 510(k): 91-2109 K061342. 91-2110 K051942. The Malibu" System is intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. The System is comprised of single use pedicle screws of varying reduction height, rods, rod connectors, etc. and an accompanying set of reusable surgical instruments. Some of the surgical instruments are specific to the system and others are general surgical instruments. The polyaxial screwdriver is specific to the polyaxial screws in the system.

Reason

Integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None of the complaints reported a patient injury. However, some complaints did indicate the surgery was slightly prolonged while the additional driver provided in the kit was retrieved and used.

Identifiers

code_info
Model Number(s): 91-2109 (8.0 Plus) 91-2110 (Standard). Lot AL549784F: 91-2109 Malibu Polyaxial Screwdr…Model Number(s): 91-2109 (8.0 Plus) 91-2110 (Standard). Lot AL549784F: 91-2109 Malibu Polyaxial Screwdriver 8.0 Plus. Lot AL552404F: 91-2110 Malibu Polyaxial Screwdriver Standard. Lot AL546768F: 91-2110 Malibu Polyaxial Screwdriver Standard. The devices are not distributed as sterile and do not have a expiration date.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0938-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.