RecallRadar

FDA Device recall Z-0929-2022

Atellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536

On 2022-04-20 the FDA classified a Class II recall of Atellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536 by Siemens Healthcare Diagnostics, citing siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0929-2022
ClassificationClass II
Statusongoing
Initiated2022-02-09
Published2022-04-20
Last updated2026-09-13 11:56 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Atellica CH Enzymatic Hemoglobin A1c (A1c_E/A1c_H), Siemens Material Number 11097536

Reason

Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Identifiers

code_info
UDI: 00630414220505; Lot Numbers: 110027 110060 110065 110091 110113 110129 110137 110161 110170 110208 110224 110234 1…UDI: 00630414220505; Lot Numbers: 110027 110060 110065 110091 110113 110129 110137 110161 110170 110208 110224 110234 110235 110262 110265 110273 110286 110315 110330 110347 110360

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0929-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.