FDA Device recall Z-0927-2020
da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and accessories to perform…
- FDA Device
- Class II
- terminated
- Published 2020-02-12
On 2020-02-12 the FDA classified a Class II recall of da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and accessories to perform… by Intuitive Surgical, citing the firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0927-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-07-26 |
| Published | 2020-02-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Reason
The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Identifiers
| code_info | System Serial # PSC Serial # VSC Serial # SP0047 Not Affected 675054 SP0049 Not Affected 674261 S…System Serial # PSC Serial # VSC Serial # SP0047 Not Affected 675054 SP0049 Not Affected 674261 SP0050 Not Affected 675051 SP0054 Not Affected 682482 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0927-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.