FDA Device recall Z-0926-2020
da Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used with an endoscope, surgical instruments, and accesso…
- FDA Device
- Class II
- terminated
- Published 2020-02-12
On 2020-02-12 the FDA classified a Class II recall of da Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used with an endoscope, surgical instruments, and accesso… by Intuitive Surgical, citing the firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0926-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-07-26 |
| Published | 2020-02-12 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Intuitive Surgical |
| Distribution | US |
Product
da Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Reason
The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
Identifiers
| code_info | System Serial # PSC Serial # VSC Serial # SL0350 665033 Not Affected SL0360 292201 Not Affected SL036…System Serial # PSC Serial # VSC Serial # SL0350 665033 Not Affected SL0360 292201 Not Affected SL0361 Not Affected 1000216 SL0362 381021 678362 SL0365 34608 Not Affected SL0366 Not Affected 1000217 SL0369 Not Affected 679520 SL0370 Not Affected 678840 SL0371 Not Affected 305493 SL0372 415418 678699 SL0373 429524 334972 SL0375 332395 678509 SL0376 Not Affected 680721 SL0377 Not Affected 991369 SL0378 266528 872469 SL0379 311710 676421 SL0380 Not Affected 268852 SL0385 Not Affected 304487 SL0386 62043 681143 SL0387 308179 687814 SL0388 Not Affected 688819 SL0389 71796 308181 SL0391 518359 285588 SL0392 Not Affected 685499 SL0393 295941 687813 SL0394 281121 687293 SL0395 43875 689242 SL0396 Not Affected 685077 SL0397 892731 685500 SL0398 Not Affected 330718 SL0399 Not Affected 685080 S |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0926-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.