RecallRadar

FDA Device recall Z-0925-2024

PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416

On 2024-02-07 the FDA classified a Class II recall of PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416 by Howmedica Osteonics, citing product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material p….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0925-2024
ClassificationClass II
Statusongoing
Initiated2023-12-01
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyHowmedica Osteonics
DistributionUS

Product

PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416

Reason

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Identifiers

code_infoGTIN: 04546540324122 Lot Numbers: 571630, 582549, 584840

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.