FDA Device recall Z-0924-2017
Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional…
- FDA Device
- Class II
- terminated
- Published 2017-02-01
On 2017-02-01 the FDA classified a Class II recall of Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional… by Philips Electronics North America, citing due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0924-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-12-01 |
| Published | 2017-02-01 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Philips Electronics North America |
| Distribution | WA |
Product
Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology(EP).
Reason
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Identifiers
| code_info | 1000 |
|---|