FDA Device recall Z-0923-2026

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

On 2025-12-31 the FDA classified a Class II recall of Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800 by Agfa Healthcare, citing it is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation wh….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0923-2026
ClassificationClass II
Statusongoing
Initiated2025-11-18
Published2025-12-31
Last updated2026-09-13 11:57 UTC
CompanyAgfa Healthcare
DistributionUS

Product

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Reason

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Identifiers

code_infoN/A

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.