RecallRadar

FDA Device recall Z-0923-2020

Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M12249…

On 2020-02-12 the FDA classified a Class II recall of Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M12249… by Lumitex, citing there is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with ex….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0923-2020
ClassificationClass II
Statusterminated
Initiated2019-12-11
Published2020-02-12
Last updated2026-09-13 11:55 UTC
CompanyLumitex
DistributionWI

Product

Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001

Reason

There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.

Identifiers

code_info
2014 Production Serial Numbers: All Serial numbers beginning with QAAT; 2015 Production Serial Numbers: All Serial nu…2014 Production Serial Numbers: All Serial numbers beginning with QAAT; 2015 Production Serial Numbers: All Serial numbers beginning with QAAU; 2016 Production Serial Numbers: All Serial numbers beginning with QAAV; 2017 Production Serial Numbers: All Serial numbers beginning with QAAW; 2018 Production Serial Numbers: Serial numbers QAAT60001 through QAAT60156

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0923-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.