FDA Device recall Z-0918-2015
Cable remote control JUPITER U, Material no.4500119, Serial number: to (incl.) 102306915
- FDA Device
- Class II
- terminated
- Published 2015-01-14
On 2015-01-14 the FDA classified a Class II recall of Cable remote control JUPITER U, Material no.4500119, Serial number: to (incl.) 102306915 by Trumpf Medical Systems, citing complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0918-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-12-03 |
| Published | 2015-01-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Trumpf Medical Systems |
| Distribution | AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI |
Product
Cable remote control JUPITER U, Material no.4500119, Serial number: to (incl.) 102306915. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.
Reason
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
Identifiers
| code_info | The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER S…The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188. |
|---|