Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation

On 2014-02-12 the FDA classified a Class II recall of Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation by Edwards Lifesciences, citing edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0918-2014
ClassificationClass II
Statusterminated
Initiated2014-01-20
Published2014-02-12
Last updated2026-09-13 11:53 UTC
CompanyEdwards Lifesciences
DistributionUS

Product

Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.

Reason

Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.

Identifiers

code_infolot 59617777 and 59658495.