RecallRadar

FDA Device recall Z-0915-2026

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

On 2025-12-24 the FDA classified a Class I recall of GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S by Ge Medical Systems China, citing potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event o….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0915-2026
ClassificationClass I
Statusongoing
Initiated2025-11-14
Published2025-12-24
Last updated2026-09-13 11:57 UTC
CompanyGe Medical Systems China
DistributionUS

Product

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

Reason

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Identifiers

code_infoFRU 2076139-001-S, distributed between June 2, 2025 and October 21, 2025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0915-2026%22