FDA Device recall Z-0909-2021
Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-…
- FDA Device
- Class II
- ongoing
- Published 2021-02-03
On 2021-02-03 the FDA classified a Class II recall of Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-… by Respironics California, citing potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0909-2021 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-11-02 |
| Published | 2021-02-03 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Respironics California |
| Distribution | US |
Product
Philips Respironics V60 Ventilator Part Number R1053618 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Reason
Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.
Identifiers
| code_info | Model: V60 Part Number: R1053618 V60 Ventilator, Japan Opt: CFLEX,AVAPS Serial Numbers: 100039155 100054624 100054738 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0909-2021%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.