RecallRadar

FDA Device recall Z-0907-2023

VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 080085400…

On 2023-01-18 the FDA classified a Class II recall of VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 080085400… by Ventana Medical Systems, citing internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0907-2023
ClassificationClass II
Statusongoing
Initiated2022-11-18
Published2023-01-18
Last updated2026-09-13 11:56 UTC
CompanyVentana Medical Systems
DistributionUS

Product

VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001

Reason

Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.

Identifiers

code_info
UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) As…UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G27114 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G31438 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H03958 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H13049 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H35110 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J12769 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J18775 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J25361 UDI-DI Code: 07613336106709 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 F31427 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G06715 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G16831 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G24564 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H00001 VENTAN

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0907-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.