FDA Device recall Z-0905-2015
Femoral Impactor ATTUNE INTUITION Impactor
- FDA Device
- Class II
- terminated
- Published 2015-01-14
On 2015-01-14 the FDA classified a Class II recall of Femoral Impactor ATTUNE INTUITION Impactor by Depuy Orthopaedics, citing ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0905-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-10 |
| Published | 2015-01-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Depuy Orthopaedics |
| Distribution | US |
Product
Femoral Impactor ATTUNE INTUITION Impactor. The ATTUNE INTUITION Impactors are re-useable instruments utilized in knee replacement procedures. The three affected impactors are used at various times within the surgical flow, each impactor is used to impact and drive home its relevant mating part: Fixed Tibial bearing, (254401003) Rotating Tibial Platform, (254401004) and Femoral impactor, (254401006).
Reason
ATTUNE INTUITION Impactors (cat. nos. 254401003, 254401004, and 254401006) have broken and produced small pieces. Should a fracture occur and not be observed during surgery, there is the potential for these small fractured pieces of the instrument to be left in the patient.
Identifiers
| code_info | CATALOG NO. : ATTUNE INTUITION Impactors Femoral Impactor, (254401006) Barcode GTIN Number: Femoral Impactor, 10603295130222 ALL LOTS |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0905-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.