RecallRadar

FDA Device recall Z-0903-2022

Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-0…

On 2022-04-20 the FDA classified a Class II recall of Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-0… by Merit Medical Systems, citing A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0903-2022
ClassificationClass II
Statusongoing
Initiated2022-03-01
Published2022-04-20
Last updated2026-09-13 11:56 UTC
CompanyMerit Medical Systems
DistributionAK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV

Product

Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F, K08-YP085A, K08-YS017B, K08-YS055A, K09-YS029A, K09-YS040F, K12-YP002A, K12-YP030A, K12-YP060A, K12-YP061A, K12-YP063C, K12-YP072, K12-YS018A, K12-YS019A, K12-YS026A, K05-01140D, K05-01706C***, K05-02595, MAP112, K05-00564L, K12-YP002A,

Reason

A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.

Identifiers

code_info
Catalog Code/Lots: 580775001/I2219771, I2278741, I2226790; IN8112/C/H2292357, H2287440, H2281404, H2298626, H2305047,…Catalog Code/Lots: 580775001/I2219771, I2278741, I2226790; IN8112/C/H2292357, H2287440, H2281404, H2298626, H2305047, H2305048, H2309356; K05-00053H/H2305217; K05-00109G/H2304656; K05-00137K/H2304653; K05-00410G/H2305216; K05-00438G/H2309130; K05-00502F/H2309122; K05-00587D/H2309211; K05-01052/H2304607; K05-01706C/H2309133; K05-01880K/H2306962; K05-02130/H2304639; K05-02595/H2304614; K09-12599A/H2295450; K10-05487P/H2279857; K12-00264/H2285040; MAP111/F/I2259670, I2279884, I2259657, I2289610; MAP112/B/P1908667; MAP112/F/I2259661, I2279915, I2279894, I2176725, I2215955, I2289611, I2279895, I2289613, I2289612; K08-YP085A/H2249051, H2302394; K08-YS017B/H2249052; K08-YS055A/H2275990; K09-YS029A/H2266470, H2302474; K09-YS040F/H2286601; K12-YP002A/H2269321, H2278279, H2283742, H2295661; K12-YP030A/H2263653; K12-YP060A/H2272013; K12-YP061A/H2249088; K12-YP063C/H2272382, H2295556; K12-YP072/H2266224; K12-YS018A/H2284956; K12-YS019A/H2302506; K12-YS026A/H2276014, H2302507; K05-01140D/H2304611

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0903-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.