VITROS Immunodiagnostic Products Intact PTH Reagent Pack: 1) UPN 10758750006287 / Product Code 6802892 - Reagent Pack, 2) UPN 1075875000625…

On 2017-01-04 the FDA classified a Class II recall of VITROS Immunodiagnostic Products Intact PTH Reagent Pack: 1) UPN 10758750006287 / Product Code 6802892 - Reagent Pack, 2) UPN 1075875000625… by Ortho Clinical Diagnostics, citing the device defect is a positive bias of up to 20% relative to the currently marketed Roche device. In addition, the firm states that during the investigation they discovered that….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0903-2017
ClassificationClass II
Statusterminated
Initiated2016-10-05
Published2017-01-04
Last updated2026-09-13 11:54 UTC
CompanyOrtho Clinical Diagnostics
DistributionUS

Product

VITROS Immunodiagnostic Products Intact PTH Reagent Pack: 1) UPN 10758750006287 / Product Code 6802892 - Reagent Pack, 2) UPN 10758750006250 / Product Code 6802893 - Calibrators For the quantitative measurement of intact parathyroid hormone (iPTH) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. Intact PTH is indicated to aid in the differential diagnosis of hyperparathyroidism, hypoparathyroidism, or hypercalcemia of malignancy and can be used intraoperatively

Reason

The device defect is a positive bias of up to 20% relative to the currently marketed Roche device. In addition, the firm states that during the investigation they discovered that the cleared reference interval (RI) was not set optimally at product launch in 2010 and that the upper limit in the IFU of 53.5pg/mL is too low and should be closer to 80 pg/mL. They are in the process of submitting the data to support the revised RI in a new 510K submission and will be changing the IFU once the 510K is cleared.

Identifiers

code_info
Lot / Expiry date: 1) 0700 (18 Oct 2016), 0710 (14 Nov 2016), 0748 (16 Feb 2017, 0758 (06 Mar 2017); 2) 0728 (12 Dec 20…Lot / Expiry date: 1) 0700 (18 Oct 2016), 0710 (14 Nov 2016), 0748 (16 Feb 2017, 0758 (06 Mar 2017); 2) 0728 (12 Dec 2016), 0738 (02 Jan 2017), 0768 (10 Apr 2017)