HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32J

On 2024-02-07 the FDA classified a Class II recall of HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32J by Howmedica Osteonics, citing product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material p….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0902-2024
ClassificationClass II
Statusongoing
Initiated2023-12-01
Published2024-02-07
Last updated2026-09-13 11:57 UTC
CompanyHowmedica Osteonics
DistributionUS

Product

HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32J

Reason

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Identifiers

code_infoGTIN: n/a Lot Numbers: 29518001

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0902-2024%22