ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

On 2015-01-14 the FDA classified a Class II recall of ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) by Horiba Instruments, citing HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0902-2015
ClassificationClass II
Statusterminated
Initiated2014-11-13
Published2015-01-14
Last updated2026-09-13 11:53 UTC
CompanyHoriba Instruments
DistributionUS

Product

ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

Reason

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

Identifiers

code_infoModel No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0902-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.