FDA Device recall Z-0902-2015
ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)
- FDA Device
- Class II
- terminated
- Published 2015-01-14
On 2015-01-14 the FDA classified a Class II recall of ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) by Horiba Instruments, citing HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0902-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2014-11-13 |
| Published | 2015-01-14 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Horiba Instruments |
| Distribution | US |
Product
ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)
Reason
HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes
Identifiers
| code_info | Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014. |
|---|