RecallRadar

FDA Device recall Z-0900-2022

LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model) [#101447…

On 2022-04-20 the FDA classified a Class II recall of LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model) [#101447… by Kaneka America, citing labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0900-2022
ClassificationClass II
Statusterminated
Initiated2022-02-17
Published2022-04-20
Last updated2026-09-13 11:56 UTC
CompanyKaneka America
DistributionAL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI

Product

LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model) [#101447, KP-05 Sulflux Plasma Separator], [#101473, HDE KP-05 Sulflux Plasma Separator], [#101785, LA-15 (AU) LDL Adsorption Column (Luer Lock Type)], [#101786, NK-M3R(UL) Tubing (Luer Lock Type)], [#101788, HDE LA-15(AU) LDL Adsorption Column (Luer Lock Type)], [#101789, HDE NK-M3R(UL) Tubing (Luer Lock Type)]. Associated labeling: 1) Instructions for Use (6): IFU LIPOSORBER LA-15 LDL Adsorption Columns FH, IFU SULFLUX KP-05 PLASMA SE-PARATOR FH, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FH, IFU LIPOSORBER LA-15 LDL Adsorption Columns FSGS, IFU SULFLUX KP-05 PLASMA SE-PARATOR FSGS, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FSGS, 2) Operators Manuals for the following (2): LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FH, LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FSGS 3) Patients Guide (1): A PATIENT GUIDE TO THE LIPOSORBER LA-15 SYSTEM FSGS

Reason

Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.

Identifiers

code_infoAll product lot and UDI numbers are affected.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0900-2022%22