RecallRadar

FDA Device recall Z-0896-2013

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700)

On 2013-03-06 the FDA classified a Class II recall of Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700) by Orthofix, citing there is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during fin….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0896-2013
ClassificationClass II
Statusterminated
Initiated2013-02-01
Published2013-03-06
Last updated2026-09-13 11:53 UTC
CompanyOrthofix
DistributionUS

Product

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Reason

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.

Identifiers

code_infoCatalog Numbers: 52-6800 & 52-6700; All Lot Numbers